next up previous contents
Next: Experimental Determination of Up: Experimental Determination of Previous: The Measured Asymmetry

Data Analysis

To determine the low temperature behaviour of $\lambda _{ab} (T)$ several assumptions were made in the fitting procedure to reduce the number of independent variables. To start with, the Ginzburg-Landau parameter $\kappa = \lambda_{ab} / \xi$ was assumed to be independent of temperature. Although this is strictly valid only for weak coupling s-wave superconductors away from Tc, the lineshapes are not very sensitive to $\kappa $ in the low-field region being considered here. Determining a value for $\kappa $was accomplished by first fitting the recorded asymmetry spectra with $\kappa $ as a fixed quantity. The value of $\kappa $ which minimized the sum of the $\chi ^{2}$'s for each temperature considered was then taken to be the best value for $\kappa $. The value $\kappa = 68$ gave the best overall fit to both the 0.5T and 1.5T data. Increasing $\kappa $ to 73 was found to change $\lambda_{ab} (0)$ by less than 0.3nm. The value $\kappa = 68$ is close to the value $\kappa = 69(1.4)$ determined from previous lineshape measurements on similar crystals in higher magnetic fields [77].

Fits to the early part of the signal ( i.e. the first $1 \mu$s) for data below Tc using an equation in the form of Eq. (4.1) with a single gaussian relaxation function of the form $\exp (- \sigma^{2} t^{2} /2)$ and $\nu_{\mu}$ pertaining to the average internal field, provides a simplified visual display of the dependence of the lineshape width on temperature. It is straightforward to use the polarization function in Eq. (4.1) to relate the relaxation parameter $\sigma$ to the second moment $\langle \Delta B^{2} \rangle$. The relationship is [68]

\begin{displaymath}\langle \Delta B^{2} \rangle = \frac{\sigma^{2}}{\gamma_{\mu}^{2}}
\end{displaymath} (4)

Comparing to Eqs. (3.6), (3.7) and (3.8), one has

\begin{displaymath}\sigma \propto \frac{1}{\lambda_{ab}^{2}}
\end{displaymath} (5)

Because of the asymmetric shape of the true field distribution, using a gaussian distribution of internal fields gives poor fits to the data. However, the fits are sufficient to provide a crude estimate of the temperature dependence of the second moment. Fig. 4.6 shows the variation of the broadening parameter $\sigma$ for single gaussian fits of the 0.5T and 1.5T data. Both sets of data suggest that the width of the field distribution varies linearly with temperature below 20K. Furthermore, the broadening of the lineshape appears larger at these low temperatures for the applied field of 0.5T. The linear term appears to weaken slightly at 1.5T. Above 20K the data for the two fields are nearly identical. The single gaussian fits of course cannot determine $\lambda _{ab} (T)$explicitly, but are useful nonetheless in giving an approximate temperature dependence of $\lambda _{ab} (T)$. They also help facilitate a comparison with previous studies where only single gaussian fits were possible.


  
Figure 4.6: The gaussian linewidth parameter $\sigma$in $\mbox{YBa$_{2}$ Cu$_{3}$ O$_{6.95}$ }$, in magnetic fields 0.5T of (triangles) and 1.5T (squares).

\begin{figure}\begin{center}\mbox{
\epsfig{file=gaussfit.eps,height=5in}
} \end{center}\end{figure}

A more precise treatment of the data using the phenomenological model of Eq. (3.39), holds the two parameters $\sigma_{eff}$ and $1/ \lambda _{ab}^{2}$ accountable for the width of the measured field distribution in the sample. Clearly these two parameters must combine to mimic the behaviour in Fig. 4.6. Since $\sigma_{eff}$ and $1/ \lambda _{ab}^{2}$ both contribute to the linewidth and both are expected to be temperature dependent quantities, the two parameters cannot be treated as independent of one another when analyzing the data. Indeed, fits to the data in which both parameters were free to vary have $\sigma_{eff}$ and $1/ \lambda _{ab}^{2}$ playing off one another as in Fig. 4.7. A temperature point which appears locally high in the $\sigma_{eff}$ vs. T plot, appears locally low in the $1/ \lambda _{ab}^{2}$ vs. T plot and vice versa. A plot of $1/ \lambda _{ab}^{2}$ vs. $\sigma_{eff}$ suggests a linear correlation between the two parameters as shown in Fig. 4.8. The solid curve through the points in Fig. 4.8 has the following form:

 \begin{displaymath}\frac{1}{\lambda_{ab}^{2}} = \frac{\sqrt{\sigma_{eff}^{2} -
\sigma_{ns}^{2}}}{C}
\end{displaymath} (6)

where the relaxation parameter $\sigma_{ns}$ [see Eq. (4.1)] is determined by a run taken above Tc in the normal state and C is a constant ( i.e. the slope) chosen so as to minimize the total $\chi ^{2}$ for all runs in which $T \leq 55$K. In other words, only runs where $T \leq 55$K were considered in determining C since it is the low temperature regime which is of primary importance.


  
Figure 4.7: The temperature dependence of (a) the gaussian linewidth parameter $\sigma_{eff}$ and (b) the magnetic penetration depth $\lambda_{ab}$in the low temperature regime at a field of 0.5T, as deduced from fitting data with independent $\sigma_{eff}$ and $\lambda_{ab}$.



  
Figure 4.8: The relationship between $\lambda_{ab}$ and the gaussian broadening parameter $\sigma_{eff}$. The solid curve is the equation which appears to the left of the data points.

\begin{figure}\begin{center}\mbox{
\epsfig{file=correlation.eps,height=4in}
} \end{center}\end{figure}


  
Figure 4.9: The total $\chi ^{2}$ for fits to the (a) 0.5T and (b) 1.5T data below 55K. The dashed lines are guides to the eye.

\begin{figure}\begin{center}\mbox{
\epsfig{file=totchi.eps,height=4in}
} \end{center}\end{figure}

Figure 4.9 shows the total $\chi ^{2}$ arising from global fits of the 0.5T and 1.5T data for various choices of the constant C. The proportionality constant C was found to be $0.0293(10) \mu \mbox{m}^{2} \mu \mbox{s}^{-1}$ and $0.0258(10) \mu \mbox{m}^{2} \mu \mbox{s}^{-1}$ for the 0.5T and 1.5T data, respectively. The depolarization rate $\sigma_{ns}$ was approximately $0.13 \mu \mbox{s}^{-1}$ and $0.11 \mu \mbox{s}^{-1}$for the 0.5T and 1.5T fields, respectively.

In the first type of analysis, the total asymmetry amplitude $A^{\circ}$ for signals recorded below Tc was fixed to the value of the precession amplitude $A^{\circ}_{ns}$ obtained from fitting data above the transition temperature, prior to determining C in Eq. (4.6). Below Tc the asymmetry amplitude of the measured signal $A^{\circ}$ is the sum of the precession amplitude of the background signal ( $A^{\circ}_{bkg}$) and the precession amplitude of the signal originating from within the sample ( $A^{\circ}_{sam}$). Thus, here we are assuming that the total precession amplitude of the resultant signal is independent of temperature, but dependent upon the applied magnetic field. The asymmetry amplitude above Tcat fields of 0.5T and 1.5T were found to be $A^{\circ} \approx 0.266(1)$ and $A^{\circ} \approx 0.247(1)$, respectively. The field dependence is primarily attributed to the finite timing resolution of the counters, which causes the observed precession amplitude to decrease as the period of the muon precession becomes comparable to the timing resolution.

In the final step of this analysis, the status of the fitting parameters was then as follows:
1. Sample Signal [refer to Eq. (3.39)]:

Variable parameters:
i) The amplitude $A^{\circ}_{sam}$
ii) $1/ \lambda _{ab}^{2}$
iii) The average internal field $\overline{B}$
iv) The initial phase $\theta$

Fixed parameters:
i) $\kappa $
ii) $\sigma_{eff}$ fixed to $\lambda_{ab}$ according to Eq. (4.6)

2. Background Signal [refer to Eq. (4.3)]:

Variable parameters:
i) The field $\overline{B}_{bkg}\approx B_{applied}$
ii) $\sigma_{bkg}$
iii) The initial phase $\theta$ (same as for sample signal)

Fixed parameters:
i) The amplitude, $A^{\circ}_{bkg}= A^{\circ}_{ns} -
A^{\circ}_{sam}$

Thus in the final fit of the data there were six independent parameters. Figure 4.10 shows the variation with temperature of the initial phase $\theta$, the average field $\overline{B}_{bkg}$,the amplitude $A^{\circ}_{bkg}$and relaxation rate $\sigma_{bkg}$ of the background precession signal, obtained from fits of the 0.5T data. As indicated in Fig. 4.10(a), the phase of the initial muon spin polarization remains nearly constant throughout the temperature scan ( i.e. $\delta \theta \sim 0.05$rad). This implies that there were no appreciable fluctuations in the applied field or electronics. The nearly constant field $\overline{B}_{bkg}$in Fig. 4.10(b) is a further indication of a highly stable applied magnetic field. The 1.5T data is not shown because there was a significant change in the applied field after 40K.

Figure 4.10(d) shows a significant drop in the relaxation rate of the background signal at higher temperatures, indicating some temperature dependence for $\sigma_{bkg}$. However, at lower temperatures ($T \leq 50$K) the background relaxation rate and hence the contribution of the background signal to the second moment exhibits no obvious correlation with temperature. This suggests that $\sigma_{bkg}$plays little role in the temperature dependence of $\sigma$ in Fig. 4.6. The fact that $\sigma_{bkg}> \sigma_{ns}$ suggests that the background is caused by a material with a large nuclear dipolar interaction such as Cu, or is in a region of fairly large field inhomogeneity.


  
Figure 4.10: The temperature dependence of (a) the initial phase $\theta$, (b) the average field $\overline{B}_{bkg}$, (c) the amplitude $A^{\circ}_{bkg}$ and (d) the depolarization rate $\sigma_{bkg}$ corresponding to the background signal produced by muons missing the sample, in an applied field of 0.5T. These results are taken from fits in which the total muon spin precession signal amplitude was fixed to a constant.

\begin{figure}\begin{center}\mbox{
\epsfig{file=background.eps,height=6.0in}
} \end{center}\end{figure}

Figure 4.11 and Fig. 4.12 show the temperature dependence of $A^{\circ}_{sam}$, $\overline{B}$ and $1/ \lambda _{ab}^{2}$ arising from the same fits which produced the results in Fig. 4.10. Together these parameters constitute three of the four variable parameters (the other being $\theta$) which pertain to the signal originating from the sample. The sample amplitude $A^{\circ}_{sam}$ depicted in Fig. 4.11(a) shows some scatter and a slight decrease at higher temperatures. The scatter in the asymmetry amplitude is not all that surprising considering that the data was recorded over a period of 5 days, through which time, small fluctuations in experimental conditions were unavoidable. For instance, one such experimental variation was the rate at which 4He was pumped through the cryostat. At higher temperatures (where the required cooling power is low) the amount of 4He flowing into the cryostat and the corresponding pumping rate were minimized in an effort to keep the heater voltage small to preserve the supply of 4He and to reduce thermal gradients between the thermometers and the sample. However, to maintain low temperatures a much larger flow of 4He was required. The increased density of helium atoms in the cryostat increases the probability of scattering the incoming muons before they can reach the sample, thus increasing the background signal and decreasing the magnitude of $A^{\circ}_{sam}$. To minimize this effect, the cryostat sample space was pumped on hard, but the choice of a specific combination of 4He-flow rate and the pumping rate was purely judgemental. This is a possible explanation for the scatter observed in Fig. 4.11(a). However, the downward trend of $A^{\circ}_{sam}$ as one increases the temperature may be purely statistical, as a similar behaviour was not observed in more recently recorded data fitted with the same procedure. Recall that since the total asymmetry amplitude was fixed, the variation of $A^{\circ}_{bkg}$ with temperature in Fig. 4.10(c) appears as a mirror image of Fig. 4.11(a).


  
Figure 4.11: The temperature dependence of (a) the muon precession amplitude $A^{\circ}_{sam}$ and (b) the average internal field $\overline{B}$ (circles), corresponding to the precession signal produced by muons hitting the sample, and the background field $\overline{B}_{bkg}$ (triangles), for an applied field of 0.5T. The total precession signal amplitude was assumed constant in the fits.

\begin{figure}\begin{center}\mbox{
\epsfig{file=sample.eps,height=4in}
} \end{center}\end{figure}

Figure 4.11(b) shows the temperature variation of the average internal field $\overline{B}$ experienced by muons implanted in the $\mbox{YBa$_{2}$ Cu$_{3}$ O$_{6.95}$ }$ sample. For comparison, the background field $\overline{B}_{bkg}$is also plotted in Fig. 4.11(b). In general, the field at any point in the sample is the sum of the local fields in Eq. (3.13). For all temperatures, $\overline{B}$is less than $\overline{B}_{bkg}$, but $\overline{B}$ appears to approach $\overline{B}_{bkg}$ at both ends of the temperature scan. In the high-temperature regime the vortex cores begin to overlap with the internal field distribution approaching full penetration of the applied field. Thus it is not surprising to see the average internal field $\overline{B}$ approach $\overline{B}_{bkg}$as one increases the temperature towards Tc. The rise in average field $\overline{B}$ at low temperatures, however, is more difficult to understand. Such an increase has also been reported in previous work by Riseman [77] and observed in more recent data taken at different fields. The cause for such behaviour is puzzling indeed. However, Fig. 4.11(b) is consistent with the time spectrum shown in Fig. 4.3 which shows a more distinct beat in the muon spin precession signal at the intermediate temperature T=35.5K, corresponding to a greater separation between the average precession frequency of muons subjected to the internal field distribution and the average precession frequency of muons in the background field. This suggests that the increase in $\overline{B}$ at low temperatures may be due to some intrinsic phenomenon of the $\mbox{YBa$_{2}$ Cu$_{3}$ O$_{6.95}$ }$sample itself.

  
Figure 4.12: The temperature dependence of $1/ \lambda _{ab}^{2}$ in a magnetic field of 0.5T, as determined from fits in which the total precession signal amplitude was assumed constant.


Figure 4.12 shows the temperature dependence of $1/ \lambda _{ab}^{2}$ (which in the phenomenological London Model is directly proportional to the superfluid density ns) for the applied field of 0.5T . Since the relaxation rate $\sigma_{eff}$ is assumed proportional to $1/ \lambda _{ab}^{2}$ [see Eq. (4.6)], the variation of $\sigma_{eff}$ with temperature resembles the behaviour in Fig. 4.12. Figure 4.13 shows the low-temperature dependence of $1/ \lambda _{ab}^{2}$for both 0.5T and 1.5T applied fields. As shown, the presence of a linear term ( i.e. $1/ \lambda_{ab}^{2} \propto T$) in the low-temperature region is unmistakeable for both 0.5T and 1.5T fields, with the latter showing a weaker linear dependence on T. A fit to the low-temperature data ( i.e. below 55K), with an equation of the form:

 \begin{displaymath}\frac{1}{\lambda_{ab}^{2}(T)} = \frac{1}{\lambda_{ab}^{2}(0)} \left[
1 - \alpha T - \beta T^{2} \right]
\end{displaymath} (7)

gives $\lambda_{ab} (0) = 1451(2) $Å and $\lambda_{ab} (0) = 1496(1) $Å for the 0.5T and 1.5T data respectively, where the quoted uncertainties are purely statistical [87]. It should be noted that this equation is purely phenomenological and cannot be extended to include the higher-temperature data. Both curves suggest that pair breaking persists at the lowest of temperatures in $\mbox{YBa$_{2}$ Cu$_{3}$ O$_{6.95}$ }$, which is inconsistent with conventional s-wave pairing of carriers. This low-temperature behaviour indicates a more complicated gap function $\Delta (\vec{k}, T)$ characteristic of the presence of nodes in the energy gap.


  
Figure 4.13: The temperature dependence of $1/ \lambda _{ab}^{2}$ in magnetic fields of 0.5T and 1.5T, as determined from fits in which the total precession signal amplitude was assumed constant. The solid lines are fits to Eq. (4.7).

\begin{figure}\begin{center}\mbox{
\epsfig{file=both1st.eps,height=4.5in}
} \end{center}\end{figure}

In Fig. 4.14 the temperature dependence of $\lambda_{ab}$ at 0.5T and 1.5T is shown. The solid curves represent microwave measurements of the change in penetration depth $\Delta \lambda_{ab} (T)$taken in zero static magnetic field by Hardy et al. [6]. For the purpose of comparison, $\lambda_{ab} (0)$ for each field is chosen to be the value obtained from fitting the $\mu ^{+}$SR low-temperature data with Eq. (4.7). Surprisingly, the microwave data shows a much better agreement with the $\mu ^{+}$SR data at the higher magnetic field of 1.5T.


  
Figure 4.14: The temperature dependence of $\lambda_{ab}$ at (a) 0.5T and (b) 1.5T. The solid lines show the microwave measurements of $\Delta \lambda_{ab} (T)$ in zero field from Ref. [6], assuming $\lambda_{ab} (0) = 1451$ and 1496Å in (a) and (b) respectively.

\begin{figure}\begin{center}\mbox{
\epsfig{file=micro1.eps,height=4in}
} \end{center}\end{figure}

There was some concern after completion of the above analysis that fixing the total asymmetry amplitude to the value above Tcmay introduce systematic errors by constraining the fits. The large fluctuation in the amplitude of the muon spin precession signal originating from the sample [see Fig. 4.11(a)] was the source of such concerns. Intuitively, we expect that $A^{\circ}_{sam}$ should scale with the percentage of muons striking the sample. The fluctuations in this percentage during the actual experiment were probably not large enough to account for the large scatter in Fig. 4.11(a). If one dismisses the previous explanation for the large fluctuations in $A^{\circ}_{sam}$, it is worth investigating this matter further.

Since $A^{\circ}_{sam}$ is not expected to change significantly over the temperature scan, the data was refitted first by designating $A^{\circ}_{sam}$ and $A^{\circ}_{bkg}$ as variable parameters. As in the previous analysis, $\sigma_{eff}$ was assumed to be proportional to $1/ \lambda _{ab}^{2}$ through Eq. (4.6). The proportionality constant C was determined to be 0.0250(10) and $0.0243(10) \mu \mbox{m}^{2} \mu \mbox{s}^{-1}$ for the 0.5T and 1.5T data respectively. The variable parameters pertaining to the background precession signal varied with temperature according to Fig. 4.15. Comparing with Fig. 4.10, the phase $\theta$ shifts down $\sim 0.005$rad, while $\overline{B}_{bkg}$ shifts upward $\sim 0.05$mT. The degree of fluctuation in both these parameters appears similar to that of the previous analysis, so again it seems apparent that there were negligible fluctuations in the applied field.


  
Figure 4.15: The temperature dependence of (a) the initial phase $\theta$, (b) the average field $\overline{B}_{bkg}$, (c) the amplitude $A^{\circ}_{bkg}$ and (d) the depolarization rate $\sigma_{bkg}$ corresponding to the background signal produced by muons missing the sample, in an applied field of 0.5T. These results are taken from fits in which $A^{\circ}_{sam}$ and $A^{\circ}_{bkg}$ were free to vary.

\begin{figure}\begin{center}\mbox{
\epsfig{file=backgroundf.eps,height=6.0in}
} \end{center}\end{figure}

The amplitude $A^{\circ}_{bkg}$ and the relaxation rate $\sigma_{bkg}$[see Fig. 4.15(c) and Fig. 4.15(d)] show almost no change from the results depicted in Fig. 4.10. Even the size of the statistical error bars are comparable. These results indicate that the fitting program is capable of clearly separating the unwanted background signal from the sample signal.

Figure 4.16(a) shows the temperature dependence of the amplitude $A^{\circ}_{sam}$ corresponding to the muon spin precession signal originating from the sample. The downward trend with increasing T appears slightly more prominent than in Fig. 4.11(a). The temperature dependence of $\overline{B}$in Fig. 4.16(b) is significantly different from that in the previous analysis. The average field $\overline{B}$ is greater than $\overline{B}_{bkg}$ at the lowest of temperatures and does not dip as far below $\overline{B}_{bkg}$ as in Fig. 4.11(b) for temperatures beyond this. At the high-temperature end in Fig. 4.11(b), $\overline{B}$ recovers to approximately the same value as in Fig. 4.11(b). Again the rise in $\overline{B}$ at low temperatures is surprising. It is possible that this is an effect due to $\vec{a}$-$\vec{b}$ anisotropy. The presence of $\vec{a}$-$\vec{b}$anisotropy would distort the vortex lattice into isoceles triangles as shown in Fig. 3.9. If this lattice were to be modelled by one consisting of equilateral triangles as assumed in our analysis, then there would be some error in the determination of the average field $\overline{B}$. This would be a greater problem at low temperatures where the cores are further apart and errors in spectral weighting are more pronounced.


  
Figure 4.16: The temperature dependence of (a) the muon precession amplitude $A^{\circ}_{sam}$ and (b) the average internal field $\overline{B}$ (circles), corresponding to the precession signal produced by muons hitting the sample, and the background field $\overline{B}_{bkg}$ (triangles), for an applied field of 0.5T. $A^{\circ}_{sam}$ and $A^{\circ}_{bkg}$ were both independently variable parameters in the fits.

\begin{figure}\begin{center}\mbox{
\epsfig{file=samplef.eps,height=4in}
} \end{center}\end{figure}

The low-temperature dependence of $1/ \lambda _{ab}^{2}$ is shown in Fig. 4.17. Surprisingly, the scatter in the data points is not significantly greater than in Fig. 4.13. Noticeably different however, is an increase in the linear term (see Table 4.1). Furthermore, fits to Eq. (4.6) yield $\lambda_{ab} (0) = 1350(2) $Å and $\lambda_{ab} (0) = 1437(1) $Å for the 0.5T and 1.5T data, respectively. A comparison to the microwave measurements of Hardy et al., assuming these values for $\lambda_{ab} (0)$ is shown in Fig. 4.18. There appears to be even less agreement at 0.5T than previously noted in Fig. 4.14(a). However, the agreement at 1.5T in Fig. 4.18(b) is comparable to that in Fig. 4.14(b) despite the significant difference in $\lambda_{ab} (0)$.


  
Figure 4.17: The temperature dependence of $1/ \lambda _{ab}^{2}$ in magnetic fields of 0.5T and 1.5T, as determined from fits in which $A^{\circ}_{sam}$ and $A^{\circ}_{bkg}$ were variable parameters. The solid lines are fits to Eq. (4.7).

\begin{figure}\begin{center}\mbox{
\epsfig{file=bothfree.eps,height=4.5in}
} \end{center}\end{figure}


  
Figure 4.18: The temperature dependence of $\lambda_{ab}$ at (a) 0.5T and (b) 1.5T. The solid lines show the microwave measurements of $\Delta \lambda_{ab} (T)$ in zero field from Ref.[6], assuming $\lambda_{ab} (0) = 1350$ and 1437Å in (a) and (b), respectively.

\begin{figure}\begin{center}\mbox{
\epsfig{file=microf.eps,height=4in}
} \end{center}\end{figure}

As a final step in the analysis, the data was refitted with the amplitude $A^{\circ}_{sam}$ fixed to the average value of the data below 55K in Fig. 4.16(a). This results in a noticeable reduction in the scatter for the parameters $A^{\circ}_{bkg}$ and $\sigma_{bkg}$ (see Fig. 4.19). Fixing $A^{\circ}_{sam}$ in this way significantly shifts the data points above 40K. This is not surprising since $A^{\circ}_{sam}$ was fixed to the low-temperature average. The phase $\theta$ shows a slight decrease at high temperatures [see Fig. 4.19(a)] and $\overline{B}$ levels off above 40K [see Fig. 4.20(b)]. These results suggest that fixing $A^{\circ}_{sam}$ to the low-temperature average reduces the scatter in the low-temperature data, but it is not yet clear whether or not we are introducing non-physical deviations in the high-temperature region.


  
Figure 4.19: The temperature dependence of (a) the initial phase $\theta$, (b) the average field $\overline{B}_{bkg}$, (c) the amplitude $A^{\circ}_{bkg}$ and (d) the depolarization rate $\sigma_{bkg}$ corresponding to the background signal produced by muons missing the sample, in an applied field of 0.5T, from fits where $A^{\circ}_{sam}$ is assumed constant.

\begin{figure}\begin{center}\mbox{
\epsfig{file=background2.eps,height=6.0in}
} \end{center}\end{figure}


  
Figure 4.20: The temperature dependence of (a) the muon precession amplitude $A^{\circ}_{sam}$ and (b) the average internal field $\overline{B}$ (circles), corresponding to the precession signal produced by muons hitting the sample, and the background field $\overline{B}_{bkg}$ (triangles), for an applied field of 0.5T.

\begin{figure}\begin{center}\mbox{
\epsfig{file=sample2.eps,height=4in}
} \end{center}\end{figure}

The reduction in scatter is most noticeable in Fig. 4.21 which shows the temperature dependence of $1/ \lambda _{ab}^{2}$. From Eq. (4.6) we find $\lambda_{ab} (0) = 1362(2) $Å and $\lambda_{ab} (0) = 1445(1) $Å for the 0.5T and 1.5T data, respectively. A plot of the temperature dependence of $1/ \lambda _{ab}^{2}$over the full temperature scan is shown in Fig. 4.22. The two fields appear to converge well before Tc, but the crossover is difficult to determine. As shown in Fig. 4.23, there is improved agreement between the microwave measurements and the 0.5T $\mu ^{+}$SR data, while the agreement with the 1.5T data is comparable to that of the previous two fitting methods. The total asymmetry amplitude of the muon spin precession signal as determined from all three fitting procedures is shown in Fig. 4.24. It appears as though one is justified in fixing the total asymmetry amplitude, as the average values are comparable.


  
Figure 4.21: The temperature dependence of $1/ \lambda _{ab}^{2}$ in magnetic fields of 0.5T and 1.5T, as determined from fits in which $A^{\circ}_{sam}$ was assumed to be constant. The solid lines are fits to Eq. (4.7).

\begin{figure}\begin{center}\mbox{
\epsfig{file=both2nd.eps,height=4.5in}
} \end{center}\end{figure}


  
Figure 4.22: The temperature dependence of $1/ \lambda _{ab}^{2}$ in magnetic fields of 0.5T and 1.5T, as determined from fits in which $A^{\circ}_{sam}$ was assumed to be constant.



  
Figure 4.23: The temperature dependence of $\lambda_{ab}$ at (a) 0.5T and (b) 1.5T. The solid lines show the microwave measurements of $\Delta \lambda_{ab} (T)$ in zero field from Ref.[6], assuming $\lambda_{ab} (0) = 1362 $ and 1445Å in (a) and (b) respectively.

\begin{figure}\begin{center}\mbox{
\epsfig{file=micro2.eps,height=4in}
} \end{center}\end{figure}


  
Figure 4.24: The temperature dependence of the total asymmetry amplitude at 0.5T for three different fitting procedures: (a) $(A^{\circ}_{sam}+ A^{\circ}_{bkg})$ is fixed to a constant value; (b) $A^{\circ}_{sam}$ and $A^{\circ}_{bkg}$ are independent parameters; (c) $A^{\circ}_{sam}$ is a fixed parameter.

\begin{figure}\begin{center}\mbox{
\epsfig{file=totasy.eps,height=3.5in}
} \end{center}\end{figure}

The results from all three types of analysis are summarized in Table 4.1. Methods (ii) and (iii) give comparable results, but differ substantially from method (i). The difference appears to be related to the proportionality constant C of Eq. (


next up previous contents
Next: Experimental Determination of Up: Experimental Determination of Previous: The Measured Asymmetry
Jess H. Brewer
2001-09-28