![]() |
The T1 relaxation is defined as the relaxation of the spin
component parallel to the external magnetic field [34]. In
order to measure the T1 relaxation with SR, one uses the
configuration shown in Fig.14a, because
the muon spin is parallel to the beam axis by default (see
Fig.6).
With the same counter configuration (Fig.14b), it is
also possible to measure T2 relaxation, which is the relaxation
of the spin components perpendicular to the external field
[34]. In this counter geometry, there is a certain upper
limit for the transverse field (
200 G for surface
muons), because the muon trajectory curves in the magnetic field, and in
the worst case, it misses the sample. For T2 relaxation
measurements in higher fields, the `Left-Right' (or `Up-Down')
configuration has to be employed (Fig.14c). The
measurement with this configuration requires a good DC-separator
on the beamline, which is capable of rotating the muon spin perpendicular
to the beam (see inset of Fig.10).
One benefit to the SR method is that measurements in
zero magnetic field are possible. This condition yields the highest
sensitivity to small internal magnetic fields. To understand the spin
relaxation in the zero-field, the spin relaxation theories developed for
the Nuclear Magnetic Resonance (NMR) method become inadequate, because
those theories assume the existence of an external magnetic field.
The next chapter introduces spin relaxation theories which are applicable to
this zero-field condition.